Keysight N2862B, N2863B, 

N2889A, and N2890A Probes

User’s Guide

Summary of Contents for N2862B

Page 1: ...Keysight N2862B N2863B N2889A and N2890A Probes User s Guide...

Page 2: ...ard com mercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www keysight com find sweula The license set forth in the EULA represents the excl...

Page 3: ...obe Shipment 4 Oscilloscope Compatibility 4 Cleaning the Probe 4 Handling the Probe 4 Switching the Attenuation Ratio on the N2889A Probe 5 Typical Voltage Derating Curve 6 Characteristics 8 Accessori...

Page 4: ...signs of stress notify the carrier as well as your Keysight Technologies Sales Office Keep the shipping materials for the carrier s inspection The Keysight Technologies office will arrange for repair...

Page 5: ...n in the picture below CAUTION Note that the probe cable is a sensitive part of the probe and therefore you should be careful not to damage it through excessive bending or pulling You should also avoi...

Page 6: ...B N2889A and N2890A User s Guide Typical Voltage Derating Curve Typical Voltage Derating Curve Figure 1 N2862B N2863B N2889A N2890A Voltage Derating Curve 300Vrms Figure 2 N2889A Voltage Derating Curv...

Page 7: ...Typical Voltage Derating Curve N2862B N2963B N2889A and N2890A User s Guide 7 Note that the maximum input voltage rating of the probe decreases as the frequency of the applied signal increases...

Page 8: ...MHz Risetime 10 90 2 33 ns 1 16 ns 1 ns 10 1 35 ns 1 1 700 ps Attenuation ratio 10 1 10 1 1 1 10 1 switchable 10 1 Input resistance when terminated into 1 M 10 M 10 M 10 M 10 1 1 M 1 1 10 M Input capa...

Page 9: ...on Usage Quantity 1 Retractable hook To make quick connections for hands free probing 1 2 Adjustment tool To adjust the LF compensation to an optimum square wave response 1 3 Ground lead black 12 cm T...

Page 10: ...D tags 2 of each color 7 Ground spring To improve measurement performance by providing a short ground connection 1 8 Insulating cap To fit over the probe tip and cover the grounding band of the probe...

Page 11: ...ible through the cable jacket Using a product with a worn cable may result in electric shock fire or equipment failure WARNING Indoor Use Only Do not operate in wet damp environments Keep product surf...

Page 12: ...onnect the probe to the oscilloscope and connect the ground lead to earth ground before connecting the probe to the circuit under test Disconnect the probe input and the probe ground lead from the cir...

Page 13: ...N2963B N2889A and N2890A User s Guide 13 Returning the Probe for Service For all repair service or calibration needs you can send the probe to an authorized service center Visit http www keysight com...

Page 14: ...tacting Keysight Technologies For technical assistance contact your local Keysight Call Center In the Americas call 1 800 829 4444 In other regions visit http www keysight com find assist Before retur...