Diagnostics and troubleshooting
Universal Data Manager
44
Diagnost
ic code
Message text
Description
Remedy
F301
Error: Cannot load
setup
Setup defective
Switch the device off and then on again,
re-configure, contact the Service
Department if necessary
M302
Setup restored from
backup
Setup has been loaded from
backup.
Check setup
F303
Error: Device data
Device data defective
Contact the Service Department
M304
Backup: Device data Device data defective. However it
was possible to continue working
with the backup data.
Check settings (e.g. serial number)
F307
Error: Customer
preset value
defective
Customer preset value defective
F309
Error: Date/time is
not set
Invalid date/time (e.g. internal
battery is empty)
Device was switched off too long. The
date/time must be set again. Battery
might need to be replaced (contact the
Service Department).
F310
Error: Cannot save
setup
The setup could not be saved.
Contact the Service Department
F311
Error: Device data
The device data could not be saved. Contact the Service Department
F312
Error: Calibration
data defective
The calibration data could not be
saved.
Contact the Service Department
F312
Analog card x is not
calibrated!
Analog card x is not calibrated! The
device works with default values,
i.e. the measured values could be
inaccurate under certain
circumstances.
Contact the Service Department
M313
SRAM has been
defragmented
SRAM was defragmented after
firmware update
No action required. Message can be
acknowledged.
F314
Error: Option code
Activation code is no longer correct
(incorrect serial number/program
name). Option has been switched
off and setup preset has been
performed.
Enter new code
M315
No IP address could
be obtained from
the DHCP server!
No IP address could be obtained
from the DHCP server!
Check the network cable
M316
Invalid MAC
address!
No or incorrect MAC address
Contact the Service Department
M317
Battery voltage < 2
V. Please replace
battery!
Battery needs to be replaced (contact
the Service Department)
F348
Firmware cannot be
updated:
• Checksum
incorrect
• Firmware
incompatible!
Firmware update has been aborted
because the firmware file is
damaged or is not compatible with
this device
Contact the Service Department
M350
Measured value
acquisition
interrupted for
calibration/service
work.
Measured value
acquisition
restarted.
Measured value acquisition was
interrupted/reactivated for service/
maintenance purposes.
Causes include:
• Calibration of inputs/outputs
• Firmware update
No action required. Message can be
acknowledged.