CELL-DYN
3000 System Operator’s Manual
10-41
9140240E — May 1995
Chapter 10
Troubleshooting
2.
, press [CNT
RATE SUMMARY].
3.
Press [WBC CNT RATE] to display the data for the kinetic rate
followed by [PRINT] to obtain a printout.
4.
Press [WBC RATE GRAPH] to display the graph of the kinetic
data followed by [PRINT] to obtain a printout.
5.
For a list of probable causes and corrective actions, refer to
Table 10.3, Sample-Related Fault Conditions
Troubleshooting Imprecise or Inaccurate Data
1.
Obtain a normal fresh whole blood sample. Select an empty QC file
and run a minimum of 10 Open mode runs into the file. Obtain a
printout of the file.
2.
Run a minimum of nine Closed mode runs into the same file. (For
the CELL-DYN 3000SL, aliquot a sample into three tubes and run
each tube three times.) Use the [REJECT SPECIMEN] key to
reject the Open mode runs and obtain a printout. This information
can be used to determine if the problem is mode or measurement
related.
3.
Compare the CVs of the precision data from a normal fresh whole
blood sample to the CVs of
Precision of the Hemogram Parameters (N = 20)
4.
Obtain printouts of related data as indicated in the following steps.
5.
WBC and Differential:
•
to display the following and obtain
printouts by pressing the Print Screen key on the keyboard.
(This requires two printouts per sample.)
Printout 1:
Size/Complexity scatterplot
Depol/Orthogonal scatterplot
0
°
/90
°
scatterplot
90
°
/10
°
scatterplot
Printout 2:
MP histogram
LBM histogram
Search