10-68
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting
Chapter 10
Table 10.3:
Sample-Related Fault Conditions
The following abbreviations are used in this table:
BL — Bulletin Line
SB — Status Box
Information for each message is listed in order from the most likely
cause of the message to the least likely cause of the message. Therefore,
always troubleshoot a problem in this order.
__________________________________________________________
Sampling Error — Incomplete Aspiration/BL
SAMPLING ERR is displayed on the
MCHC result.
SAMPLING ERR is printed on the graphics and pre-printed ticket report.
SAMPLING ERROR is printed on the blank ticket.
– – – – – Probable Cause – – – –
– – – Corrective Action – – – – –
1.
The blood sensors did
1.
Check the sample tube to
not detect a sufficient
be sure it contains a
amount of sample
sufficient quantity of during
aspiration.
blood.
2.
section of this chapter or
the
remove any obstructions.
3.
4.
5.
Call for technical assistance.