![Abbott CELL-DYN 3000 Operator'S Manual Download Page 459](http://html1.mh-extra.com/html/abbott/cell-dyn-3000/cell-dyn-3000_operators-manual_450973459.webp)
10-30
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting
Chapter 10
Unclogging the Open Sample Aspiration Probe
The open sample aspiration probe is thoroughly cleaned whenever the
Auto-Clean procedure is performed. If a blockage is suspected, it may be
cleared as directed in the following procedure.
Materials Required
1.
Monofilament line
2.
Empty Vacutainer® tube
3.
Enzymatic Cleaner
4.
Gloves, lab coat, and suitable eye protection
WARNING: Potential Biohazard. Follow established
biosafety practices when performing maintenance, service or
troubleshooting procedures. Wear gloves, lab coat, and suitable
eye protection. The probe is potentially contaminated with
infectious material. Refer to
Chapter 8, Precautions, Limitations
CAUTION: Use powder-free gloves when performing this
procedure to avoid contaminating the instrument.
Procedure
1.
Remove the tubing from the top of the open sample aspiration
probe as indicated in
NOTE: The tubing is connected with a silicon tubing sleeve.
Disconnect the sleeve from the probe.
Search
Go Back
Book TOC