Chapter 6
Calibration
Table 6.5:
Calibration Criteria Chart
Determining Which Parameters Need Calibration
1.
Validation Range
If the Factor % Difference is equal to or less than the following
values, CALIBRATION IS NOT REQUIRED for that parameter as
the value is within the Validation Range.
Results EQUAL TO OR LESS THAN VALIDATION RANGE:
WBC
<1
.5%
RBC
<1
.0%
HGB
<1
.0%
MCV
<1
.0%
PLT
<
3.0%
If desired, the manual calibration results may be documented in the
is available for comments only when
a calibration factor is changed or reentered. Therefore, document
the calibration in the <COMMENTS> line by accessing the log as
follows:
press [ENTER
FACTOR].
Retype any existing Cal Factor. Press the Enter key on the keyboard
to save the entry and advance the cursor.
Press [RETURN].
the <COMMENTS> line that instrument calibration was
confirmed and no changes were required. Press the Enter key on
the keyboard to save the entry and advance the cursor.
Validation Range
Cal Not Required
Calibration Range
Cal Required
Calibration Limit
Do Not Cal
WBC
<1.5%
>1.5% but
<
10%
>10%
RBC
<1.0%
>1.0% but
<
10%
>10%
HGB
<1.0%
>1.0% but
<
10%
>10%
MCV
<1.0%
>1.0% but
<
10%
>10%
PLT
<3.0%
>3.0% but
<
15%
>15%