6-88
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
* Current factor as printed in this manual calibration procedure.
** If factor exceeds limits, do not calibrate. Check all calculations and call the Customer Support Center for assistance at 1 (800)
CELL DYN.
* Mean after calibration.
** If % Diff exceeds limits, call the Customer Support Center for assistance at 1 (800) CELL DYN.
New Closed Mode Calibration Factors
Open Mode Mean x Current Closed Mode Cal Factor* = New Closed Mode Cal Factor
Closed Mode Mean
Open Mode
Mean
/
Closed
Mode
Mean
x
Current
Closed Mode
Cal Factor*
=
New
Closed
Mode Cal
Factor
Range**
WBC
/
x
=
0.700 - 1.300
RBC
/
x
=
0.800 - 1.200
HGB
/
x
=
0.700 - 1.300
MCV
/
x
=
0.700 - 1.300
PLT
/
x
=
0.700 - 1.300
Mode to Mode Post-Calibration Difference
Closed Mode Mean - Open Mode Mean x 100 = % Diff
Open Mode Mean
Closed
Mode
Mean*
-
Open
Mode
Mean
/
Open Mode
Mean
x 100 =
%Diff
Range**
WBC
-
/
x 100 =
<1
.75%
RBC
-
/
x 100 =
<1
.25%
HGB
-
/
x 100 =
<1
.25%
MCV
-
/
x 100 =
<1
.25%
PLT
-
/
x 100 =
<
3.50%