9-16
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Maintenance
Chapter 9
Closed Sampler Aspiration Needle
The aspiration needle on the closed sampler version of the CELL-DYN
3000 should be cleaned regularly to remove protein buildup or debris and
reduce the possibility of a blockage.
Materials Required
1.
CELL-DYN Enzymatic Cleaner (brought to room temperature)
2.
Diluent
3.
Three empty Vacutainer® tubes
4.
Gloves, lab coat, and suitable eye protection
Procedure
1.
Aliquot approximately 2 mL of enzymatic cleaner (undiluted) into
one of the Vacutainer tubes.
2.
Aliquot approximately 2 mL of diluent into each of the other two
Vacutainer tubes.
3.
, press [CLOSED
SAMPLER] to select the closed mode.
4.
Press [SPECIMEN TYPE] followed by [BACKGROUND].
5.
Aspirate the enzymatic cleaner.
6.
Aspirate the Diluent.
7.
5 background counts on the last diluent sample. If the counts are
unacceptable, refer to
8.
Discard all tubes used.