Chapter 6
Calibration
Calibrating the Open Mode
1.
, press [CALIBRATION].
2.
Press [ENTER FACTOR] to display the
3.
Type the New Open Mode Calibration Factors for the parameters that
require calibration into the appropriate fields in the <OPEN
SAMPLER> column. Move the cursor to the desired position, type
the factor and press the Enter key on the keyboard to save the factor
and advance the cursor.
4.
After the New Open Mode Calibration Factors have been entered,
press [RETURN].
5.
The log holds 10 entries. (The screen displays five entries. Display
the other five by pressing the Page Up key on the keyboard.) When
the log is full, subsequent entries cause the oldest entry to be
deleted and the remaining entries to move up one line, so that the
current factors are added to the bottom of the list. Therefore, the
log should be printed periodically for purposes of documentation.
The log displays the <DATE>, <TIME>, <OPERATOR ID>,
<CALIBRATION FACTORS> and a line for <COMMENTS>.
NOTE: The letters in parentheses after each factor indicate the
method of factor derivation: A = Autocal, F = Factory,
E = Enter Factor (manual factor entry).
6.
Type any comments in the <COMMENTS> field. Press the Enter key
on the keyboard to save the entry and advance the cursor.
NOTE: Comments may be added to the Calibration Log only
after a calibration factor is changed or reentered.
7.
If desired, press [PRINT LOG] to obtain a printout of the
Calibration Log for documentation.
Completing Manual Calibration
1.
Press [RETURN] to return to the
2.
Press [MAIN] to exit the
3.
Retain all necessary documentation in the instrument logbook.
4.
Confirm the calibration of the Open mode by running commercial
controls as directed in the