Troubleshooting
Chapter 10
Table 10.1
Instrument Status Conditions (Continued)
Auto Sampler cannot be started if the safety cover is off/BL
– – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – –
The Sample Loader START key was pressed and the safety interlock
switch did not sense that the safety cover was in place.
Replace the safety cover and then press the START key.
__________________________________________________________
Auto Sampler Emergency stop/BL
– – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – –
The Sample Loader E-STOP key was pressed during operation and the
Sample Loader is stopped.
Press the INIT key (if the light on the key is blinking) to initialize the
Sample Loader. If the INIT key light is not blinking, turn the Sample
Loader power switch OFF and ON to initialize.