CELL-DYN
3000 System Operator’s Manual
10-37
9140240E — May 1995
Chapter 10
Troubleshooting
Syringes
If a syringe is cracked or broken, it should be replaced. It may also be
necessary to replace a syringe if no other solution to a problem is found.
If this action is indicated by the Customer Support Center, obtain the
appropriate replacement syringe. Then, use the following procedure:
1.
Remove the syringe in question as directed in the
2.
Install the new syringe as directed in the syringe cleaning
procedure and follow the remaining steps in that procedure.
3.
Repeat samples that were run when the problem was detected to
determine if it has been resolved. If the problem is not resolved,
call for technical assistance.