CELL-DYN
3000 System Operator’s Manual
6-81
9140240E — May 1995
Chapter 6
Calibration
Manual Mode To Mode Calibration
Procedure
The calibration procedure has been divided into subsections that consist
of a series of easy-to-follow steps. Always follow the entire procedure
unless specifically directed to skip to another section.
Confirm the calibration of the Open mode before performing this
procedure.
NOTE: Be sure to complete the
sections before beginning this procedure.
Preparing for Manual Mode to Mode Calibration
1.
, press [CALIBRATION].
2.
If necessary, press [CLOSED SAMPLER] to display the Closed
Mode Calibration Factors.
3.
Press [PRINT] to obtain a printout of the <CURRENT WHOLE
BLOOD FACTORS> displayed on the screen.
4.
Press [RETURN] followed by [MAIN] to return to the
Determining the Open Mode Mean
1.
, press [RUN].
2.
If necessary, press [CHANGE SAMPLER] to select the Open
mode.
3.
Press [SPECIMEN TYPE].
4.
Use the arrow keys on the keyboard to move the cursor to an empty
QC file and type “OPEN MEAN” in the <FILENAME> field.
5.
Press the Enter key on the keyboard to save the name. Use the Up
Arrow key on the keyboard to move the cursor back to the “OPEN
MEAN” file.
6.
Press [QC SPECIMEN] to return to the
7.
Run each sample one time into the “OPEN MEAN” QC file.
8.
Press [MAIN] followed by [QC LOGS] and [VIEW QC LOG].
9.
With the cursor at "OPEN MEAN" file, press [PRINT QC LOG]
to obtain a printout of the data.