6-22
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
The [REPEAT SPECIMEN] key (see Figure 6.9) is displayed when the
first run of the first specimen is completed. This key is used to delete all
results (runs) for the current specimen.
NOTE: This key deletes all results for all runs of the current
specimen. It does not delete one run only.
Figure 6.9:
Whole Blood Auto-Cal Results Screen
The [CALIBRATR] key is used to display the
). This screen accesses the Auto-Cal
program that is used to calibrate the instrument with a commercial
calibrator. The following soft key labels are displayed:
EDIT REF VAL
START AUTO-CAL
QUIT AUTO CAL
CLEAR REF VALS
PRINT SUMMARY
RETURN
OPEN
SAMPLER
REPEAT
SPECIMEN
REPEAT
SPECIMEN
INTERRUPT
AUTO-CAL
QUIT
AUTO-CAL
RETURN
WHOLE BLOOD AUTO-CAL
Ready for Calibration
May 11 1992
Operator ID
Sequence #
10:30
734
4100
Open Sampler
RESULTS FOR SPECIMEN #1
Spec 1
Mean
Factor
Parameter
Value
RUN1
RUN2
RUN3
Factor
Factor
%Diff
WBC
7.30
7.42
----
----
0.980
----
----
RBC
4.24
4.39
----
----
0.970
----
----
HGB
12.9
13.0
----
----
0.990
----
----
MCV
90.0
91.0
----
----
0.990
----
----
PLT
244.
254.
----
----
0.960
----
----
SPEC ID CAL#01-01
NO. OF RUNS 3
CALIBRATR