1-14
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
System Description
Chapter 1
Sheath Reservoir
The Sheath reservoir maintains the Sheath reagent supply within the
Analyzer.
Hemoglobin Lyse Reservoir
The Hemoglobin Lyse reservoir maintains the Hemoglobin Lyse supply
within the Analyzer.
Normally Closed Valves
The three normally closed valves prevent the reagents in the reservoirs
from draining down into the Analyzer when the Analyzer power is turned
OFF.
Analyzer Serial Number Label
This label contains the manufacturer's serial number for the Analyzer.
Analyzer Power Switch
This is the main power switch for the Analyzer. It is used to power the
instrument ON and OFF.
Auto Sampler Connector
The Auto Sampler Connector port is used to attach the serial interface
cable from the optional Sample Loader.
Waste Sensor Connector
The Waste-Full sensor plug connects to the waste sensor connector port.
When the electrical sensor is tripped, the WASTE FULL message is
generated and the READY status is inhibited until the situation is
corrected. The Analyzer interprets a disconnected plug the same way as a
full waste container. Therefore, if the waste is routed to a drain, a dummy
plug must be inserted in the connector.
Diluent Inlet Tube Connector
This color-coded (red) port is used to connect the Diluent inlet tube with
its associated cap, sinker and label.
Sheath Inlet Tube Connector
This color-coded (purple) port is used to connect the Sheath inlet tube
with its associated cap, sinker and label.