Calibration
Chapter 6
CELL-DYN 3000
Auto-Cal Calibration Worksheet
Instrument:______________________________________
Date:________________________________________
Operator:________________________________________
________________________________________________________________________________________
_______
* Delete the sign of the Factor % Difference before entering it on the chart.
** Do not calibrate. Call the Customer Support Center for assistance at 1 (800) CELL DYN.
Auto-Cal Calibration Criteria
Factor
%Diff*
Validation Range
Cal Not Required
Cal Range
Cal Required
Cal Limit
Do Not Cal**
Cal?
Y/N
WBC
<
1.5%
>1.5% but
<
10%
>10%
RBC
<
1.0%
>1.0% but
<
10%
>10%
HGB
<
1.0%
>1.0% but
<
10%
>10%
MCV
<
1.0%
>1.0% but
<
10%
>10%
PLT
<
3.0%
>3.0% but
<
15%
>15%