CELL-DYN
3000 System Operator’s Manual
1-9
9140240E — May 1995
Chapter 1
System Description
•
HGB Diluent syringe — delivers a specific volume of Diluent to
transport the blood from the shear valve to the HGB mixing
chamber
•
HGB Lyse syringe — dispenses a specific volume of Hemoglobin
Lyse into the HGB Mixing chamber at the same time as the diluted
sample is dispensed into the HGB Mixing chamber
•
WBC Sheath syringe — delivers a specific volume of Sheath
reagent to transport the blood from the shear valve to the WBC
Mixing chamber
•
WBC Metering syringe — injects a specific volume of the WBC
dilution into the WBC flow cell
Waste Chambers
The waste chambers collect the waste liquid from the Analyzer flow
panel.