CELL-DYN
3000 System Operator’s Manual
6-57
9140240E — May 1995
Chapter 6
Calibration
* Current Open Mode factor as printed at the start of this calibration procedure.
* Delete the sign of the % Diff before entering it on the chart.
** Do not calibrate. Call the Customer Support Center for assistance at 1 (800) CELL DYN.
Factor % Difference
New Open Mode Factor - Current Open Mode Factor x 100 = % Diff
New Open Mode Factor
New Open
Mode
Factor
-
Current Open
Mode Factor*
/
New Open
Mode Factor
x 100 =
%Diff
WBC
-
/
x 100 =
RBC
-
/
x 100 =
HGB
-
/
x 100 =
MCV
-
/
x 100 =
PLT
-
/
x 100 =
Manual Calibration Criteria
%Diff*
Validation Range
Cal Not Required
Cal Range
Cal Required
Cal Limit
Do Not Cal**
Cal?
Y/N
WBC
<1.5%
>1.5% but
<
10%
>10%
RBC
<1.0%
>1.0% but
<
10%
>10%
HGB
<1.0%
>1.0% but
<
10%
>10%
MCV
<1.0%
>1.0% but
<
10%
>10%
PLT
<3.0%
>3.0% but
<
15%
>15%