9-22
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Maintenance
Chapter 9
Procedure
1.
, press [DISABLE
ANALYZER].
2.
Remove the upper front cover to gain access to the peristaltic
pumps.
3.
Locate the sample aspiration pump to the left of the shear valve.
(See
4.
Pull the release lever away from the rollers and remove the tubing
from the metal brackets that hold it on each side of the rollers. (See
5.
Disconnect the tubing at the plastic fittings.
6.
Connect the new tubing to the plastic fittings.
7.
Place the collars on the ends of the tubing into the metal guides.
Hold the release lever open and push the tubing underneath the
pump rollers. Position the tubing in the center of the rollers.
8.
Replace the upper front cover.
9.
Press [ENABLE ANALYZER].
10.
Press [MAIN] to return to the
11.
Verify that the background counts are acceptable by running 5
background counts before running controls. If the
or controls are unacceptable, refer to
Sample Loader Tray, Racks and Safety Cover
The Sample Loader tray, racks and safety cover should be cleaned on a
regular basis. Blood spills in the tray or racks should be cleaned up
immediately to allow proper movement of the racks. Weekly cleaning is
recommended but more frequent cleaning may be indicated by the
laboratory workload.
Materials Required
1.
Container (large enough to hold a rack) filled with a mild detergent
solution made with warm (not hot) water
2.
Clean, warm water for rinse
3.
Lint-free wipes or towels
4.
Gloves, lab coat, and suitable eye protection