10-34
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting
Chapter 10
Figure 10.22:
Sample Loader Vent/Aspiration Needle Assembly
Procedure
1.
the procedure is complete, turn the Sample Loader power switch
OFF.
2.
Remove all racks from the tray.
3.
Place some gauze under the aspiration needle and/or vent needle to
catch any liquid.
4.
Disconnect the sample aspiration tubing from the Analyzer.
NOTE: The tubing is connected on the left side of the open
sample aspiration probe.
Aspiration
Tubing
Locking
Mounting
Bracket
Aspiration
Needle
Aspiration
Needle
Wash Block
Waste
Tubing
Port
Vent
Tubing
Locking
Screw
Mounting
Bracket
Vent
Needle
Vent
Needle
Wash
Block
Mixing
Head
Screw