Calibration
Chapter 6
CELL-DYN 3000
Mode to Mode Calibration Worksheet
Instrument:__________________________________ Date:______________________________________
Operator:______________________________________
________________________________________________________________________________________
* Delete the sign of the Factor % Difference before entering it on the chart.
** Do not calibrate. Call the Customer Support Center for assistance at 1 (800) CELL DYN.
Mode to Mode Calibration Criteria
Factor
%Diff*
Validation Range
Cal Not Required
Cal Range
Cal Required
Cal Limit
Do Not Cal**
Cal?
Y/N
WBC
<1.75%
>1.75% but
<
10%
>10%
RBC
<1.25%
>1.25% but
<
10%
>10%
HGB
<1.25%
>1.25% but
<
10%
>10%
MCV
<1.25%
>1.25% but
<
10%
>10%
PLT
<3.50%
>3.50% but
<
20%
>20%