6-4
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
Complete instructions for using these criteria and a Calibration Criteria
Chart are included with each procedure to facilitate the decision. The
chart is depicted in Table 6.1.
Table 6.1:
Calibration Criteria Chart
The calibration procedures have been divided into subsections that
consist of a series of easy-to-follow steps. Always follow the entire
procedure unless specifically directed to skip to another section.
Worksheets are provided at the end of each procedure to assist in
determining which parameters require calibration. For manual
calibration, worksheets may be used to assist in making the necessary
calculations. These worksheets may be duplicated as needed.
NOTE: Always complete the
beginning any calibration.
Conventions Used in this Chapter
used in this chapter is given in the
Validation Range
Cal Not Required
Calibration Range
Cal Required
Calibration Limit
Do Not Cal
WBC
<1.5%
>1.5% but
<
10%
>10%
RBC
<1.0%
>1.0% but
<
10%
>10%
HGB
<1.0%
>1.0% but
<
10%
>10%
MCV
<1.0%
>1.0% but
<
10%
>10%
PLT
<3.0%
>3.0% but
<
15%
>15%