CELL-DYN
3000 System Operator’s Manual
6-73
9140240E — May 1995
Chapter 6
Calibration
•
If the first run passes this internal Reference Check, proceed to
•
If the first run fails the Reference Check for any parameter, the
results are highlighted and the bulletin line displays the
message:
THIS RESULT IS OUTSIDE THE ALLOWED LIMITS.
REPEAT THIS SPECIMEN
NOTE: Check to be sure the reference values were entered
correctly and the correct sample was run before repeating it.
The specimen may be repeated at this time by pressing [REPEAT
SPECIMEN] and [CONFIRM REPEAT]. When the repeat
function is used, all results for the failed run are automatically
deleted.
•
If the repeated run also fails the Reference Check, the result(s)
are highlighted and no calibration factor will be calculated for
that parameter.
•
If all parameters fail the Reference Check, results of that run
are not displayed and the bulletin line displays the message:
ABANDON CAL FOR SPECIMEN 1 - RESULT NOT
INCLUDED IN MEAN
Repeat the specimen as directed above.
•
If all parameters on the repeated run also fail the Reference
Check, discard that sample and continue with the next
calibration sample. If desired, a replacement sample may be
added after the remaining samples have been run.
9.
Run each sample for the number of runs entered in
” section. Each subsequent run
must pass the Reference Check and an additional internal Precision
Check. If a run fails either check, that parameter is highlighted and
no calibration factor will be calculated for the parameter.
Calibration must be repeated for the highlighted parameter.
Calibration factors will be calculated for the remaining parameters.
10.
When all runs of a sample have been completed, the bulletin line
displays the message:
NEW CAL FACTOR COMPUTED. READY FOR NEXT CAL
SPECIMEN
Press [NEXT SPECIMEN] to run the next sample.