![Abbott CELL-DYN 3000 Operator'S Manual Download Page 440](http://html1.mh-extra.com/html/abbott/cell-dyn-3000/cell-dyn-3000_operators-manual_450973440.webp)
CELL-DYN
3000 System Operator’s Manual
10-11
9140240E — May 1995
Chapter 10
Troubleshooting
Figure 10.8:
Raw Data Summary Screen
When the [MORE] key is pressed, the second
is displayed. The [MORE] keys on the remaining screens always
display the next screen. Consequently, they are discussed last in each
section.
When the [PRINT] key is pressed, a diagnostic report is printed. This
report contains information pertinent to the data displayed on the screen
at the time the key is pressed. If no data is displayed on the screen, the
report prints the current fault status.
The [PRINT] key functions in this way on each screen. Therefore, it will
not be discussed again in this section.
The [MAIN] key is used to return to the
[MAIN] key appears on each primary
and works the same way on each screen. Consequently, it will not be
discussed again in this section.
FAULT
REPORT
EXECUTION
TIMES
CNT RATE
SUMMARY
CLEAR
FAULTS
RAW DATA
SUMMARY
MORE
MAIN
DIAGNOSTICS MENU
Ready
Jan 28 1993
Operator ID
Sequence #
12:33
734
3143
List Mode WBC: 4350 RBC: 23598 PLT: 2682
Raw Count
Count Time Upper: 3.53
Count: 6.23
Avg: 6.22
Timeout: 6.51
HGB Sample
1: 1218
2: 1218
3: 1219
4: 1219
5: 1219
HGB Ref
2255
2254
2255
2255
2258
RBC Alg
RER: 33.9%
Lo thr: 40
CTRUE: 342.4
WBC Alg
% Tot: 98.6%
PLT Alg
Adj cnt: 2728
Lo thr: 8
Hi thr: 219
2741
31166
4206
MORE
MAIN
Search
Go Back
Book TOC