6-56
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
CELL-DYN 3000
Manual Calibration Worksheet
Instrument:____________________________________ Date:___________________________________
Operator:______________________________________
________________________________________________________________________________________
_
Calculate All Factors To Three Decimal Places
* Current factor as printed in this manual calibration procedure.
** If factor exceeds limits, do not calibrate. Check all calculations and call the Customer Support Center for assistance at 1 (800)
CELL DYN.
New Open Mode Calibration Factors
Assay or Ref Mean x Current Open Mode Cal Factor* = New Open Mode Cal Factor
Open Mode
Assay Value
or
Ref Mean
/
Open
Mode
Mean
x
Current
Open Mode
Cal Factor*
=
New Open
Mode Cal
Factor
Range**
WBC
/
x
=
0.700 - 1.300
RBC
/
x
=
0.800 - 1.200
HGB
/
x
=
0.700 - 1.300
MCV
/
x
=
0.700 - 1.300
PLT
/
x
=
0.700 - 1.300