10-60
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting
Chapter 10
Table 10.2
General Fault Conditions (Continued)
Flow sequence time out <x,x>/BL
NOTE: Characters in the brackets identify the problem flow
sequence number and flow sequence name.
– – – – – Probable Cause – – – –
– – – – Corrective Action – – – –
1.
An internal Analyzer flow
1.
Record the characters in the
sequence was not completed
brackets. Call the Customer
in the allotted time.
Support Center to report
the problem at
1 (800) CELL DYN.
Initialize the Analyzer by
pressing the
[INITIALIZATION] key on
the second
processing if the fault does
not recur.
__________________________________________________________
Command sent at incorrect time/BL
– – – – – Probable Cause – – – –
– – – – Corrective Action – – – –
1.
The Analyzer received a
1.
Initialize the Analyzer by
command from the Data
pressing the
Station at the incorrect
[INITIALIZATION] key on
time and could not
process it.
call for technical assistance.