![Abbott CELL-DYN 3000 Operator'S Manual Download Page 354](http://html1.mh-extra.com/html/abbott/cell-dyn-3000/cell-dyn-3000_operators-manual_450973354.webp)
CELL-DYN
3000 System Operator’s Manual
7-5
9140240E — May 1995
Chapter 7
Quality Control
Figure 7.2:
The X-B Display Screen (Data)
When the [SHOW GRAPHS] key is pressed, the
for the graphs is displayed. (See
Batch
MCV
MCH
MCHC
DATE
TIME
Batch
MCV
MCH
MCHC
DATE
TIME
1
90.36
30.34
33.70
10/17/92
13:35
11
89.84
29.52
32.79
12/10/92
10:38
2
90.23
30.34
33.82
10/19/92
13:59
12
90.16
29.50
32.80
01/12/93
10:45
3
93.23
30.84
33.34
10/22/92
10:11
13
90.69
30.16
33.32
01/18/93
13:38
4
91.42
30.73
33.43
10/26/92
15:45
14
91.91
30.90
33.37
01/18/93
15:58
5
91.44
30.74
33.62
10/30/92
10:52
15
92.74
31.12
33.37
01/21/93
11:13
6
90.64
30.76
34.59
11/02/92
15:46
16
89.29
30.01
33.48
02/01/93
11:13
7
90.64
30.72
33.82
11/16/92
10:58
17
89.07
30.03
33.82
02/02/93
14:23
8
90.62
29.39
32.33
11/17/92
09:50
18
89.08
30.04
33.82
02/04/93
14:23
9
90.61
29.56
32.59
12/07/92
10:58
19
89.66
30.12
33.70
03/08/93
15:36
10
89.54
29.50
32.78
12/08/92
09:42
20
89.07
29.81
33.68
03/09/93
10:53
Upper
94.50
31.50
35.70
Upper
94.50
31.50
35.70
Lower
85.50
28.50
32.30
Lower
85.50
28.50
32.30
SHOW
GRAPHS
RETURN
X-B DISPLAY
Uninitialized
Mar 12 1993
Operator ID
Sequence #
14:28
734
2368
SHOW
GRAPHS
Search
Go Back
Book TOC