![Abbott CELL-DYN 3000 Operator'S Manual Download Page 340](http://html1.mh-extra.com/html/abbott/cell-dyn-3000/cell-dyn-3000_operators-manual_450973340.webp)
CELL-DYN
3000 System Operator’s Manual
6-85
9140240E — May 1995
Chapter 6
Calibration
Results WITHIN THE FOLLOWING CALIBRATION RANGE:
WBC
>1.75% but
<1
0%
RBC
>1.25% but
<1
0%
HGB
>1.25% but
<1
0%
MCV
>1.25% but
<1
0%
PLT
>3.50% but
<
20%
Calibration Factor Calculation
1.
Use the appropriate worksheet to calculate the New Closed Mode
Calibration Factor for the parameters that require calibration as
follows: (Use the Current Closed Mode Cal Factor printed at the
start of this procedure.)
If the New Closed Mode Calibration Factor falls within the
acceptable range given on the worksheet, calibrate the Closed
Mode as directed in the next section.
If the New Closed Mode Calibration Factor for a given parameter
falls outside the acceptable range, there may be a computation
error. DO NOT CALIBRATE THAT PARAMETER. Recheck all
calculations and then call the Customer Support Center for
assistance at 1 (800) CELL DYN.
Calibrating the Closed Mode
1.
, press [CALIBRATION].
2.
Press [ENTER FACTOR] to display the
3.
Type the New Closed Mode Calibration Factors for the parameters
that require calibration in the appropriate fields in the <CLOSED
SAMPLER> column. Move the cursor to the desired position, type
the factor and press the Enter key on the keyboard to save the factor
and advance the cursor.
4.
After the new Closed mode calibration factors have been entered,
press [RETURN].
OpenMode Mean
Closed Mode Mean
-----------------------------------------------
Current Closed Mode
Calibration Factor
×
New Closed Mode
Calibration Factor
=
Search
Go Back
Book TOC