CELL-DYN
3000 System Operator’s Manual
10-47
9140240E — May 1995
Chapter 10
Troubleshooting
Table 10.1
Instrument Status Conditions (Continued)
Initialized/SB
– – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – –
The Analyzer hardware initialization is complete.
Press [PRIME] or [RUN] to initiate the priming cycle and return the
instrument to the READY state. Resume operation when the cycle is
complete.
__________________________________________________________
Unpinching valves/SB
– – – – – – –– – – – – – – Explanation/Action– – – – – – – – – – – – – – –
The Analyzer was idle for a predetermined time period and therefore the
valves are being exercised to be sure that tubing is not pinched shut.
Resume processing when the READY message is displayed in the status
box.
__________________________________________________________
Extended Count/SB
– – – – – – –– – – – – – – Explanation/Action– – – – – – – – – – – – – – –
A low value has been detected for the WBC and/or PLT count. The
Analyzer is automatically extending the cycle to count more cells.
Press [PRIME] to resume processing.
__________________________________________________________
Clearing RBC Orifice/SB
– – – – – – –– – – – – – – Explanation/Action– – – – – – – – – – – – – – –
The [CLEAR ORIFICE] key was pressed or the instrument
automatically performed the Clear Orifice function in response to a clog
or flow problem.
Wait until the READY message is displayed in the status box and repeat
the sample.