5-34
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Operating Instructions Chapter 5
4.
Auto-transmission of NON-ALERTED graphic data
When this option is enabled, histograms and scatterplots are
automatically transmitted to the LIS for any sample without
flagged results.
5.
The remaining options are configured according to the
requirements of the LIS:
Transmission CTS enabled
Transmission Data bits (7, 8)
Transmission Stop bits (1, 2)
Transmission Parity (0=None, 1=Odd, 2=Even)
Transmission time out (0.1 to 9.9)
Computer Baud Rate (300, 600, 1200, 2400, 4800, 9600)
The numbers in parentheses after the options indicate the selections
available.
NOTE: Refer to the Interface Specification for complete
information on interfacing or call the Customer Support Center
at 1 (800) CELL DYN.
Procedure Computer Setup
1.
From the
[COMPUTER SETUP] to display the
2.
For the first five options on the list, use the Arrow keys on the
keyboard to move the cursor to the desired selection and press
[TOGGLE ON/OFF] to enable or disable the selection.
NOTE: The [TOGGLE ON/OFF] key is displayed when the
cursor is positioned in any of the first five entry fields.
3.
For the last five options on the list, type the appropriate information
and press the Enter key on the keyboard to save the entry and
advance the cursor.
4.
When all the information has been entered, press [REINIT
INTERFACE] to initialize the interface for the selected
configuration.
5.
If desired, press the Print Screen key on the keyboard to obtain a
printout of the configuration.
6.
Press [RETURN] to return to the
7.
Press [RETURN] to return to the