Chapter 4 System Specifications
Bar Code Specifications — CELL-DYN 3000SL
Bar Code Format
The following formats with or without check digits are acceptable:
•
Code 39
•
Interleave 2 of 5
•
Codabar
Bar Code Label Specifications
Bar code labels must meet the following specifications:
•
Printed on good quality label stock
•
0.25-inch quiet zone on each end
•
0.01-inch minimum narrow bar width
•
2:1 wide to narrow bar ratio
•
0.5-inch minimum bar length
•
2-inch maximum label length
•
1.25-inch maximum label width
NOTE: Refer to