![background image](http://html1.mh-extra.com/html/abbott/cell-dyn-3000/cell-dyn-3000_operators-manual_450973113.webp)
4-2
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
System Specifications
Chapter 4
Power Specifications — CELL-DYN 3000SL
Table 4.3:
Power Specifications
Consumption
Analyzer:
900 Watts
Data Station:
300 Watts
Printer:
145 Watts
Sample Loader:
50 Watts
1395 Watts maximum (4760 BTU per hour)
Analyzer Input Requirements
Setting
Range
Frequency
100
90—110 VAC
50/60 Hz
120
110—130 VAC
50/60 Hz
220
200—240 VAC
50/60 Hz
240
220—250 VAC
50/60 Hz
Data Station Input Requirements
Setting
Range
Frequency
120
90—130 VAC
50/60 Hz
240
180—260 VAC
50/60 Hz
Printer Input Requirements
Setting
Frequency
120
50/60 Hz
Sample Loader Input Requirements
Setting
Range
Frequency
115
90—130 VAC
60 Hz
230
190—240 VAC
50 Hz