CELL-DYN
3000 System Operator’s Manual
6-87
9140240E — May 1995
Chapter 6
Calibration
CELL-DYN 3000
Manual Calibration Worksheet
Instrument:____________________________________ Date:___________________________________
Operator:______________________________________
________________________________________________________________________________________
_
Calculate All Factors To Three Decimal Places
* Delete the sign of the % Diff before entering it on the chart.
** Do not calibrate. Call the Customer Support Center for assistance at 1 (800) CELL DYN.
Mode to Mode Calibration Difference
Closed Mode Mean - Open Mode Mean x 100 = % Diff
Open Mode Mean
Closed Mode
Mean
-
Open Mode
Mean
/
Open Mode
Mean
x 100 =
% Diff
WBC
-
/
x 100 =
RBC
-
/
x 100 =
HGB
-
/
x 100 =
MCV
-
/
x 100 =
PLT
-
/
x 100 =
Mode to Mode Calibration Criteria
%Diff*
Validation Range
Cal Not Required
Cal Range
Cal Required
Cal Limit
Do Not Cal**
Cal?
Y/N
WBC
<1.75%
>1.75% but
<
10%
>10%
RBC
<1.25%
>1.25% but
<
10%
>10%
HGB
<1.25%
>1.25% but
<
10%
>10%
MCV
<1.25%
>1.25% but
<
10%
>10%
PLT
<3.50%
>3.50% but
<
20%
>20%