6-8
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Calibration
Chapter 6
3.
No more than two hours should elapse between the
CELL-DYN 3000 run and the assay by reference methodology. If
samples are run on the CELL-DYN 3000 first, assay by reference
methodology should be completed within one hour. (Certain
reference methodologies are sensitive to RBC swelling caused by
in vitro deoxygenation.)
4.
Mean values should be calculated for each parameter for each
sample from the reference assay results. These mean parameter
values can then be entered in the Auto-Cal program as reference
values for each sample.
NOTE: A worksheet is provided on the following page. This
worksheet may be used to assist with the calculation of the
parameter reference means and may be duplicated as needed.
5.
If Auto-Cal is not being used, the mean parameter values should be
averaged to obtain the cumulative mean value for each parameter.