![background image](http://html1.mh-extra.com/html/abbott/cell-dyn-3000/cell-dyn-3000_operators-manual_450973338.webp)
CELL-DYN
3000 System Operator’s Manual
6-83
9140240E — May 1995
Chapter 6
Calibration
Table 6.8:
Mode to Mode Calibration Criteria Chart
Determining Which Parameters Need Calibration
1.
Validation Range
If the % difference is equal to or less than the following values,
CALIBRATION IS NOT REQUIRED for that parameter as the
value is within the Validation Range.
Results EQUAL TO OR LESS THAN VALIDATION RANGE:
WBC
<1
.75%
RBC
<1
.25%
HGB
<1
.25%
MCV
<1
.25%
PLT
<
3.50%
If desired, the Manual Mode to Mode Calibration results may be
documented in the
The
is available for comments only when
a calibration factor is changed or reentered. Therefore, document
the calibration in the <COMMENTS> line by accessing the log as
follows:
press [ENTER
FACTOR].
Retype any existing Cal Factor. Press the Enter key on the keyboard
to save the entry and advance the cursor.
Press [RETURN].
Mode to Mode Calibration Criteria
Validation Range
Cal Not Required
Calibration Range
Cal Required
Calibration Limit
Do Not Cal
WBC
<1.75%
>1.75% but
<
10%
>10%
RBC
<1.25%
>1.25% but
<
10%
>10%
HGB
<1.25%
>1.25% but
<
10%
>10%
MCV
<1.25%
>1.25% but
<
10%
>10%
PLT
<3.50%
>3.50% but
<
20%
>20%