CELL-DYN
3000 System Operator’s Manual
6-13
9140240E — May 1995
Chapter 6
Calibration
Calibration Menu
by pressing the [CALIBRATION] key.
displays the current calibration
factors for the mode indicated, the calibration method used, the date and
time the factors were entered and the operator ID.
The following soft key labels are displayed:
ENTER FACTOR
CALIBRATN LOG
AUTO-CALIBRATE
OPEN SAMPLER
or
CLOSED SAMPLER
(Key label alternates between these
two selections when the soft key is
pressed.)
MAIN
Figure 6.1:
Calibration Menu Screen Displaying Open Mode
Calibration Factors
CALIBRA-
TION
ENTER
FACTOR
CALIBRATN
LOG
AUTO-
CALIBRATE
CLOSED
SAMPLER
MAIN
CALIBRATION
Ready
May 11 1992
Operator ID
Sequence #
10:30
734
4100
Open Sampler
Current Whole Blood Factors:
Parameter
Method
Factor
Date
Time
Operator
WBC
CALIBRATOR
1.008
04/29/92
10:38
C03
RBC
CALIBRATOR
1.071
04/29/92
10:38
C03
HGB
CALIBRATOR
1.124
04/29/92
10:38
C03
MCV
CALIBRATOR
0.985
04/29/92
10:38
C03
PLT
CALIBRATOR
1.028
04/29/92
10:38
C03
MPV
FACTORY
1.000
03/02/92
8:38
C03