CELL-DYN
3000 System Operator’s Manual
6-15
9140240E — May 1995
Chapter 6
Calibration
The [MAIN] key is used to return to the
The [ENTER FACTOR] key displays the
showing the current Open and Closed Calibration
Factors (See Figure 6.3). Calibration factors may be changed on this
screen by moving the cursor to the desired position, typing the new factor
and pressing the Enter key on the keyboard. The following soft key labels
are displayed:
RESTORE FACTORS
RESET ALL TO 1.00
RETURN
The [RESTORE FACTORS] key is used to restore the previous
calibration factors. This key is only active immediately after factors have
been changed.
Figure 6.3:
Enter Whole Blood Factor Screen
MAIN
ENTER
FACTOR
RESTORE
FACTORS
RESTORE
FACTORS
RESET ALL
TO 1.00
RETURN
ENTER WHOLE BLOOD FACTOR
Ready
May 11 1992
Operator ID
Sequence #
10:34
734
4100
Open Sampler
Closed Sampler
Parameter
(Factor Range)
Factor
Factor
WBC
(0.70..1.30)
1.008
1.056
RBC
(0.80..1.20)
1.071
1.060
HGB
(0.70..1.30)
1.124
1.136
MCV
(0.70..1.30)
0.985
0.985
PLT
(0.70..1.30)
1.028
0.939
MPV
(0.70..1.30)
1.000
1.000
Search
Go Back