7-8
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Quality Control
Chapter 7
4.
The remainder of the screen displays the file information and the
data. The upper and lower limits entered are displayed immediately
above the parameter names. The sequence number for each result is
displayed to the left of the data. The date, time and operator ID
when the sample was run are displayed to the right of the data.
5.
The following QC log codes are displayed in the column
immediately preceding the date. These codes are the same as those
used in the data log:
O — Sample was run in the open mode
C — Sample was run in the closed mode
N — Incomplete aspiration in the open mode
I — Incomplete aspiration in the closed mode
K — Metering fault (clog or flow error)
M — Mixing error on the Sample Loader
B — Blood detected in Sample Loader sample aspiration line
6.
The statistics are displayed below the data as follows:
N:
the number of runs used in the calculation
Mean:
the mean value for the number of runs used in the
calculation
Std Dev:
the standard deviation for the number of runs used in
the calculation
CV (%):
the coefficient of variation in percent for the number of
runs used in the calculation
The following soft key labels are displayed on the
(Key label alternates between
selections)
RETURN
The function of each of these keys is discussed in this section.