CELL-DYN
3000 System Operator’s Manual
10-55
9140240E — May 1995
Chapter 10
Troubleshooting
Table 10.2
General Fault Conditions (Continued)
Uninitialized/SB
– – – – Probable Cause – – – –
– – – – Corrective Action – – – –
1.
Data Station has power
1.
Ensure that the Analyzer but
the Analyzer is not
power cord is connected to
responding.
the Analyzer and that the
cord is connected to the
power outlet. Initialize the
Analyzer by pressing the
[INITIALIZATION] key on
the second
2.
Communication mal-
2.
Check the cable that function
between the
connects the Analyzer
Analyzer and Data Station.
to the Data Station. If
necessary, secure the
connections. Initialize the
Analyzer by pressing the
[INITIALIZATION] key on
the second
__________________________________________________________
Initialization Failed
– – – – – Probable Cause – – – – –
– – – – Corrective Action – – – –
1.
The Data Station was
1.
Initialize the Analyzer by
unable to initialize. The
following the instructions
CELL-DYN software does
not display the
chapter. If the Data Station
does not initialize, press
the Print Screen key on the
computer keyboard to
document any screen
messages, then call for
technical assistance.