CELL-DYN
3000 System Operator’s Manual
10-27
9140240E — May 1995
Chapter 10
Troubleshooting
Troubleshooting Procedures
The procedures in this section are for troubleshooting purposes only. A
specific procedure should be performed when indicated in this chapter or
at the request of an Abbott Customer Support Specialist.
WARNING: Potential Biohazard. Follow established
biosafety practices when performing maintenance, service or
troubleshooting procedures. Wear gloves, lab coat, and suitable
eye protection. Refer to
Chapter 8, Precautions, Limitations and
CAUTION: Use powder-free gloves when performing these
procedures to avoid contaminating the instrument.
Power ON Procedure
IMPORTANT: The laser must be allowed to warm up for 15 minutes
after the power is turned ON, if the power has been OFF more than five
minutes. Do not process samples during this warm-up period.
1.
Verify that all components are properly installed (for example,
syringes, shear valve, etc.).
2.
Verify that all reagents are properly installed.
3.
Verify that all necessary cables and power cords are properly
connected.
4.
Verify that the Analyzer covers are properly installed. If the
instrument is a CELL-DYN 3000SL, verify that the Sample Loader
safety cover is in place.
5.
Verify that the cause of the power OFF situation has been
corrected.
6.
Turn the power switches ON in the following order:
a.
Analyzer
b.
Data Station
c.
Sample Loader, if present
d.
Printer
7.
When the INITIALIZED message appears in the status box on the
, prime the system by pressing [PRIME] or
[RUN], whichever is displayed.
Search
Go Back
Book TOC