![Abbott CELL-DYN 3000 Operator'S Manual Download Page 471](http://html1.mh-extra.com/html/abbott/cell-dyn-3000/cell-dyn-3000_operators-manual_450973471.webp)
10-42
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting
Chapter 10
If necessary, refer to the directions given in
•
Obtain printouts of the WBC kinetic data for several samples.
From the first
, press [CNT
RATE SUMMARY].
Press [WBC CNT RATE] to display the data for the kinetic
rate followed by [PRINT] to obtain a printout.
Press [WBC RATE GRAPH] to display the graph of the
kinetic data followed by [PRINT] to obtain a printout.
•
If there is a flagging problem or a problem with an abnormal
sample, obtain a printout of a normal sample for comparison.
•
immediately after the problem sample is run. From the first
, press [RAW DATA
SUMMARY] followed by [PRINT] to obtain a printout.
6.
RBC, HGB, MCV and PLT:
•
histograms. Obtain printouts of problem samples.
•
immediately after the problem sample is run. From the first
, press [RAW DATA
SUMMARY] followed by [PRINT] to obtain a printout.