Chapter 9
Maintenance
Daily
Daily Maintenance Procedures
NOTE: After performing maintenance procedures, the
Auto-Clean procedure should be performed.
Auto-Clean
The Auto-Clean procedure is a fully automated procedure designed to
clean the shear valve, the WBC mixing chamber, the WBC flow cell, the
HGB mixing chamber, the HGB flow cell, the RBC/PLT transducer
assembly, the needle in the CS or SL version and all the associated fluidics.
The forward and reverse action of the peristaltic pumps is used during this
procedure to gently scrub and remove any fibrin or debris within the
system. The Auto-Clean procedure (see Figure 9.5) takes approximately
11 minutes.
Figure 9.5:
Special Protocols: Auto-Clean Screen
WARNING: Potential Biohazard. Follow established
biosafety practices when performing maintenance, service or
troubleshooting procedures. Wear gloves, lab coat, and suitable
eye protection. Refer to
Chapter 8, Precautions, Limitations and
START
CANCEL
SPECIAL PROTOCOLS
Ready
Mar 15 1993
Operator ID
Sequence #
10:09
734
2723
Hold a bottle of Cell-Dyn Enzymatic Cleaner under the probe,
and press the START key.
When the auto-clean cycle is completed,
run three or more background counts.
This cleaning process takes about 11 minutes.