CELL-DYN
3000 System Operator’s Manual
6-23
9140240E — May 1995
Chapter 6 Calibration
These soft keys have the same functions as described for the
NOTE: If Auto-Cal was last performed with a calibrator, the
following key labels are displayed when the [WHOLE
BLOOD] key is pressed:
CONFIRM SELECTION
CANCEL SELECTION
These keys are used to confirm or cancel the Whole Blood Auto-Cal
selection. The bulletin line displays the message:
PRESS CONFIRM SELECTION TO CLEAR PREVIOUS
AUTO-CAL REFERENCE VALUES.
The [LATEX] key is used to display the LATEX AUTO-CAL Screen.
This screen accesses the Auto-Cal program for MPV calibration with
latex particles which is used by Abbott service personnel.
Figure 6.10: Calibrator Auto-Cal Screen
LATEX
Open Sampler
Enter reference value for each parameter to be calibrated: (up to 10 specimens)
Spec ID # of runs WBC RBC HGB MCV PLT
(1 to 10) (4.6-10.2) (3.00-5.50) (10.0-15.5) (85.0-97.0) (150.-400.)
CAL#01 10 7.90 4.46 13.2 85.4 257.
CAL#02 3 ---- ---- ---- ---- ----
EDIT
REF VAL
START
AUTO-CAL
QUIT
AUTO-CAL
CLEAR
REF VALS
SUMMARY
RETURN
CALIBRATOR AUTO-CAL
Ready
May 11 1992
Operator ID
Sequence #
10:01
734
4100