Multi-instrument setups
R&S
®
SMM100A
462
User Manual 1179.1341.02 ─ 05
Remote command:
:SCONfiguration:EXTernal:IQOutput<ch>:RF:COUPling
:SCONfiguration:EXTernal:RF<ch>:RF:COUPling
(Delta) RF Frequency/ RF Level
In uncoupled mode, sets the RF frequency and RF level of the external instrument.
With enabled "RF Coupled" parameter, applies a frequency and level offset. The RF
frequency of the external instrument is calculated as follows:
RF
ExtInstr
= RF
Instrument
+ Delta
Freq
Both the RF frequency and the RF frequency offset are applied at the external instru-
ment.
Remote command:
:SCONfiguration:EXTernal:IQOutput<ch>:RF:FREQuency
:SCONfiguration:EXTernal:IQOutput<ch>:RF:FREQuency:OFFSet
:SCONfiguration:EXTernal:IQOutput<ch>:RF:POWer
:SCONfiguration:EXTernal:IQOutput<ch>:RF:POWer:OFFSet
:SCONfiguration:EXTernal:RF<ch>:RF:FREQuency
:SCONfiguration:EXTernal:RF<ch>:RF:FREQuency:OFFSet
:SCONfiguration:EXTernal:RF<ch>:RF:POWer
:SCONfiguration:EXTernal:RF<ch>:RF:POWer:OFFSet
RF State
With enabled "RF Coupled" parameter, displays the RF output state of the connected
external instrument. The parameter is coupled to the state of the RF output of the
R&S
SMM100A.
In uncoupled mode, enables/disables the RF output in the external instrument.
Remote command:
:SCONfiguration:EXTernal:IQOutput<ch>:RF:STATe
:SCONfiguration:EXTernal:RF<ch>:RF:STATe
9.4.2.2
External instrument configuration settings
Access:
1. Select "Taskbar > System Config > System Configuration > External RF and I/Q".
2. Navigate to the required connector.
Select "External Instruments > Config ...".
The "<Dig. Conn. Name>: External Instrument Configuration" dialog opens and dis-
plays the "External Instrument Configuration > Remote Config" settings.
System configuration settings