Remote control commands
R&S
®
SMM100A
895
User Manual 1179.1341.02 ─ 05
[:SOURce<hw>]:INPut:MODext:IMPedance<ch>
<Impedance>
Sets the impedance (50 kOhm or High = 100 kOhm to ground) for the externally sup-
plied modulation signal.
Parameters:
<Impedance>
G50 | HIGH
*RST:
HIGH
Example:
INP:MOD:IMP 50
sets 50 kOhm to ground.
Manual operation:
See
[:SOURce<hw>]:INPut:TRIGger:SLOPe
<Slope>
Sets the polarity of the active slope of an applied instrument trigger.
Parameters:
<Slope>
NEGative | POSitive
*RST:
POSitive
Manual operation:
See
13.16.8
SOURce:IQ subsystem
The
SOURce:IQ
subsystem contains the commands for configuring the I/Q modula-
tion.
Required options
See
Chapter 6.1, "Required options"
...........................................................................................895
.............................................................................................. 896
................................................................................................896
............................................................................................ 896
......................................................................................897
.................................................................................. 897
.......................................................................................... 897
[:SOURce<hw>]:IQ:SOURce
<Source>
Selects the input signal source for the I/Q modulator.
Parameters:
<Source>
BASeband | ANALog
External signals disable the amplitude modulation, an enabled
custom digital modulation, any configured digital standard or an
applied digital baseband input signal.
*RST:
BASeband
SOURce subsystem