Configuring the RF signal
R&S
®
SMM100A
307
User Manual 1179.1341.02 ─ 05
7.8.3
LO coupling settings
Access:
► Select "RF" > "LO Coupling ...".
The "Local Oscillator (LO) Coupling" settings enable you to define the coupling of
the local oscillator signal in the RF signal paths.
The related remote commands are described in
Mode
Selects the coupling mode of the LO signals.
Chapter 7.8.4, "How to configure cascaded instruments"
Note:
Working with an
external
oscillator signal disables all setting parameters of the
"RF Frequency / Phase / Ext. Reference" dialog.
In
external
, the LO coupling mode is defined within an RF
frequency range of 80
MHz to 6
GHz. If you the set the RF frequency below 80
MHz,
the R&S
SMM100A adjusts the frequency to the minimum value and displays a corre-
sponding message.
"Internal"
Uses the internal signal.
"External"
Uses an externally supplied signal.
Remote command:
[:SOURce<hw>]:FREQuency:LOSCillator:MODE
Out State
Activates or deactivates the LO output.
"On"
Switches the LO output signal on.
For coupling two instruments, the internal local oscillator signal is also
available at the LO Out connector.
"Off"
Switches the LO output signal off.
Remote command:
[:SOURce<hw>]:FREQuency:LOSCillator:OUTPut:STATe
Local oscillator coupling