Configuring the RF signal
R&S
®
SMM100A
308
User Manual 1179.1341.02 ─ 05
Mandatory LO IN Frequency
Displays the frequency that must be applied as external LO frequency.
Remote command:
[:SOURce<hw>]:FREQuency:LOSCillator:INPut:FREQuency?
LO OUT Frequency
Displays the current LO frequency at the LO Out connector.
Remote command:
[:SOURce<hw>]:FREQuency:LOSCillator:OUTPut:FREQuency?
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
A blinking LED on the front/rear panel indicates the selected connector, too (see
ter 11.2.2.4, "Status indicators"
Adjust LO Level at Current Frequency
Adjusts the level of the oscillator signal to optimum, if an external LO signal is applied.
The function is enabled in the following cases:
●
An external "LO Coupling Mode" is selected.
●
"Out State" is on.
Note:
If your test setup consists of several cascaded Rohde
&
Schwarz instruments,
always use an external oscillator signal and start the LO level adjustment with the first
instrument. Continue in the order up to the last.
Remote command:
:CALibration<hw>:LOSCillator:COUPling:LOCal?
7.8.4
How to configure cascaded instruments
This section describes how you can find out the location of the LO signal connectors
(LO In/Out). You can also find instructions for configuring the LO signals of several dis-
connected signal generators.
To find the physical Location of the LO In/Out connectors
The LO In and LO Out connectors are located on the rear panel. To find out the exact
position, proceed as follows:
1. In the block diagram, select the "T/E/P/L/V" status indication field above the "RF"
block.
2. Select "LO In" > "Show".
3. Observe the connectors's position on the "Rear" panel.
Chapter 11.2.4, "RF connectors settings"
Local oscillator coupling