![Rohde & Schwarz SMM100A Скачать руководство пользователя страница 954](http://html.mh-extra.com/html/rohde-and-schwarz/smm100a/smm100a_user-manual_1477939954.webp)
Remote control commands
R&S
®
SMM100A
954
User Manual 1179.1341.02 ─ 05
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl
<Dwell>
Sets the dwell time for a frequency sweep step.
Parameters:
<Dwell>
float
Range:
0.001 to 100
Increment: 100E-6
*RST:
0.01
Example:
See
Example "Setup an RF frequency or power sweep"
Manual operation:
See
[:SOURce<hw>]:SWEep[:FREQuency]:MODE
<Mode>
Sets the cycle mode for the frequency sweep.
Parameters:
<Mode>
AUTO | MANual | STEP
AUTO
Each trigger event triggers exactly one complete sweep.
MANual
The trigger system is not active. You can trigger every step indi-
vidually by input of the frequencies with the command
STEP
Each trigger event triggers one sweep step. The frequency
increases by the value entered with
SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic
(logarithmic spacing).
*RST:
AUTO
Example:
See
Example "Setup an RF frequency or power sweep"
Manual operation:
See
[:SOURce<hw>]:SWEep[:FREQuency]:POINts
<Points>
Sets the number of steps within the RF frequency sweep range.
See
Chapter 7.10.2.1, "Correlating parameters in sweep mode"
Two separate
POINts
values are used for linear or logarithmic sweep spacing (
SOURce<hw>]:SWEep[:FREQuency]:SPACing LIN | LOG
). The command
always affects the currently set sweep spacing.
SOURce subsystem