Remote control commands
R&S
®
SMM100A
807
User Manual 1179.1341.02 ─ 05
Example:
See
Example "Creating test signals"
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:OFFSet
<Offset>
Sets the DC component.
Parameters:
<Offset>
float
Range:
-1 to 1
Increment: 0.001
*RST:
0
Default unit: FS
Example:
Example "Creating test signals"
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:SAMPles
<Samples>
Sets the number of sample values required for the rectangular signal per period.
Parameters:
<Samples>
integer
Range:
4 to 1000
*RST:
100
Example:
See
Example "Creating test signals"
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:FREQuency
<Frequency>
Sets the frequency of the simple sinusoidal test signal.
Parameters:
<Frequency>
float
Range:
100 to depends on the installed options
Increment: 0.01
*RST:
1000
Default unit: Hz
Example:
See
Example "Creating test signals"
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:PHASe
<Phase>
Sets the phase offset of the sine wave on the Q channel relative to the sine wave on
the I channel.
SOURce subsystem