Remote control commands
R&S
®
SMM100A
784
User Manual 1179.1341.02 ─ 05
–
EGC1|EGC2
: External global clock
●
In primary-secondary instrument mode, the external baseband synchronization sig-
nal (
BBSY
)
●
OBASeband|BEXTernal|EXTernal
: Setting only
Provided only for backward compatibility with other Rohde
&
Schwarz signal gener-
ators.
The R&S
SMM100A accepts these values and maps them automatically as follows:
EXTernal = EGT1
,
BEXTernal = EGT2
,
OBASeband = INTA
Parameters:
<Source>
INTernal|EGT1|EGT2|EGC1|EGC2|EXTernal|BBSY
*RST:
INTernal
Example:
See
Example "Adjusting clock, marker and trigger settings"
Manual operation:
See
[:SOURce<hw>]:BB:DM:TRIGger:SLENgth
<SLength>
Defines the length of the signal sequence to be output in the
SINGle
trigger mode.
Parameters:
<SLength>
integer
Range:
1 to 4294967295
*RST:
1000
Default unit: symbol
Example:
See
Example "Adjusting clock, marker and trigger settings"
Manual operation:
See
[:SOURce<hw>]:BB:DM:TRIGger:RMODe?
Queries the status of signal generation.
Return values:
<RMode>
STOP | RUN
Example:
SOURce1:BB:DM:TRIGger:SOURce ELTRigger
SOURce1:BB:DM:TRIGger:SEQuence ARETrigger
SOURce1:BB:DM:TRIGger:RMODe?
Response: RUN
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:BB:DM:TRIGger:EXTernal:SYNChronize:OUTPut
<Output>
Enables signal output synchronous to the trigger event.
SOURce subsystem