Configuring the RF signal
R&S
®
SMM100A
310
User Manual 1179.1341.02 ─ 05
b) Check if both paths are able to deal with the desired RF frequency.
● If both paths can handle the RF frequency, select "RF > LO Coupling >
Mode > A
External
&
A->B
Coupled"
● If path A is not sufficiently equipped, select "RF > LO Coupling > Mode >
A
RF
Off
&
B
External"
● If path B is not able to handle the frequency, select "RF > LO Coupling >
Mode > A
External
&
B
RF
Off"
c) Set "Out State > On".
d) If the attenuation between the interconnection LO Out and LO In is more than
1
dB at the oscillator frequency, we recommend that you execute "Adjust LO
Level Current Frequency".
Note:
Complete the adjustment procedure before you perform the adjustment
on a subsequent instrument.
The signal generator automatically adjusts the internal LO signal level at the
I/Q modulator and LO Out connector.
4. Configure
Signal Generator 3.
a) Set "Freq = 5
GHz".
b) Select "RF > LO Coupling > Mode > External".
c) Set "Out State > Off".
d) If the attenuation between the interconnection LO Out and LO In is greater than
1
dB at the oscillator frequency, we recommend that you perform level adjust-
ment:
● Check if the "Adjust LO Level Current Frequency" procedure of the previ-
ous instrument is completed.
● Execute "Adjust LO Level Current Frequency".
The signal generator automatically adjusts the internal LO signal level at the
I/Q modulator and LO Out connector.
All instruments generate the RF signals in phase, as they use the same oscillator
signal looped through all instruments.
Local oscillator coupling